BET Surface Area Analyzer

BET Surface Area Analyzer

  • Item No.:TOB-S-4300
  • Order(Moq):1
  • Payment:L/C,T/T,Western Union
  • Product Origin: China
  • Shipping Port: XIAMEN
  • Lead Time:7

BET Surface Area Analyzer

1 The BET Surface Area Analyzer work principle isothermal physical international general adsorption static volumetric method, the automatic control of the whole computer, without the need for manual monitoring.

2 TOB-S4300 Series BET Surface Area Analyzer is an automatic specific surface area and pore size analyzer, which can simultaneously up to three samples are analyzed, developed by our company’s equipment operating software.

3 The instrument can be a single point, more BET than multiple data surface area, mesoporous BJH, DR and MP pore, pore distribution, pore size, and total pore volume and average pore size of the area, and the analysis of its specific surface area analysis, ranging from 0.01m2/g to no known upper bound analysis, aperture range is 0.35-500nm, which can meet the demands of various working conditions.

 

Model BET Surface Area Analyzer TOB-S-4300
Source voltage 110v-240v customized, 50/60HZ
Warranty Year limited warranty with lifetime support
Measuring Method Vacuum static volumetric method
Adsorbed gas N2,H2,He,Kr,CO,CO2,CH4, Etc.
Analysis station    3 Parallel samples analysis export

1 independent saturated vapor pressure export

 

 

Measuring range ≥0.01M2/g
Analysis range 0.35-500nm
Micropore analysis 0.35-2nm
Mesoporous analysis 2-50nm
Macropore analysis 50-500nm
Report output The BET specific surface area/analysis/adsorption isotherm

adsorption and desorption by /BJH pore volume,

pore area analysis / total pore volume / total pore area /

Langmuir method of the specific surface area analysis

 

 

Analysis accuracy ≤±1%
Pressure detection system Imports pressure sensor measurement

(0-10Torr 0-1000Torr)

Pressure detection accuracy 0.15%
Analysis temperature -196 – 100 Degree Celsius
Partial pressure range P/P0 4×10-7-0.995
Vacuum system 4 x 104Pa (German mechanical pump) or 1 x 106Pa

(German molecular pump) can choose

 Software control system INTERNET remote control
Sample pretreatment Sample independent preparation station (degassing)
Sample pipe size 10ml